Main Category:
ELECTRIC, MAGNETIC AND ELECTROMAGNETIC FIELD MEASUREMENT SENSORS
Secondary Category:
HIGH FREQUENCY
Specification
Campo Elettrico: No
Campo Magnetico: No
Campo Elettromagnetico: Sì
Status:
AVAILABLE
Brand:
NARDA PMM
Model Name:
EP-333
Item Code:
NARDA PMM EP-333 STD
Applications:
Measures Work Environments EMF
Environmental measures EMF RF
Environmental Air Quality Measures
Description:
Calibration probe for the measurement of the electrical component of the electromagnetic field in the range 100 kHz - 3.6 GHz
General Features:
Measurements shown on the certificate:
Correction factor in frequency
Linearity
Anisotropy
Calibration E2PROM
Specifications:
Flatness (with correction disabled): 0.3 - 3500 MHz = 3 dB
Flatness (with correction disabled): 3.5 - 3200 MHz = 1.5 dB
Flatness (with correction disabled): 20 - 500 MHz = 0.75 dB
Isotropicity: 0.8 dB (0.5 dB typical)
H-field rejection> 20 dB
Error at 20 ° C temperature: 60 ° C + / - 0.1 dB, 0 ° C: 20 ° C - 0.05 dB / ° C -20 ° C 0 ° C -0.15 dB / ° C
E 2 PROM internal calibration
Dimensions: length 385 mm, 133 mm diameter
Weight: 293 g
Associated Courses:
RADIO FREQUENCY ELECTROMAGNETIC FIELDS: environmental monitoring techniques
UNCERTAINTIES IN MEASURING EQUIPMENT