Main Category:
MEASUREMENT AND TEST INSTRUMENTATION
Secondary Category:
ELECTRIC, MAGNETIC AND ELECTROMAGNETIC FIELD METERS
Specification
Campo Elettrico: Sė
Campo Magnetico: Sė
Campo Elettromagnetico: Sė
Status:
AVAILABLE
Brand:
NARDA PMM
Model Name:
AMB-8057-03
Item Code:
NARDA PMM AMB-8057-03 DB
Applications:
Measures Work Environments EMF
Environmental measures EMF LF
Environmental measures EMF RF
Description:
Continuous monitoring station by a large band of electric fields, magnetic and electromagnetic fields from 10 Hz to 7 GHz (depending on the probe) with a solar panel, modem GSM
General Features:
Available in various field probes as needed. Direct sensor input connector standard Narda
Multiple 8057 units (can be connected to the base station via the GSM network) can be used to build reliable control systems and to cover large geographical areas, including national coverage.
Automatic discharge and manual data entry on the PC base station via the GSM modem internal acquisition software dedicated to data collection and analysis.
Internal rechargeable by solar panel
Dual band GSM module with SIM card holder entitled to receive and transmit data to and from landline and mobile
RS 232 or landline modem. Output parallel service for internal diagnostics.
Security Alarms: microswitch tamper switch, internal over-temperature, low battery, sondaguasta.
Dimensions: 60 x 780 (diameter x height) mm
Weight: 2.4 Kg
Internal Battery Ni-Cd 4 V 2.5 A / h rechargeable by solar panel
Specifications:
Running time: More than 6 months with the GSM stand-by for 30 min and 1 min data per day
IP54
Automatic Generation of TXT file on each dowload made
SMS: up to two phones
Automatic self-test during power up and every 7 days or remotely.
Operating temperature: -10 to + 50 ° C
Dynamic depending on the probe used
Scope of work: Electric from 0.02 V / m - 200 V / m from 50 nT Magnetic - 200 mT
Resolution: Electric 0.01 V / m; 1 nT magnetic
Sensitivity-dependent probe
Accuracy: Probe dependent
Unit of measurement: V / m; nT; mT
Measuring range: X, Y, Z and total Max, AVG, RMS
Sampling: X, Y, Z every 3 sec
Interval: 30 sec, 15 min
Maximum time of acquisition: With probe 1 and 3-band: 338 days with 15 min retention interval, 11 days with 30 sec interval memorizzazzione. Probe 4-band: 169 days with 15 min retention interval, 5 days with 30 sec interval memorizzazzione
Data Download: manual, automatic control unit operated by a set time, automatic PC
Functions: AVG, RMS, peak, recording temperature and battery voltage
Alarms: Two programmable thresholds (threshold of concern and alarm) and automatic alert of overcoming them that their return to their limits.
Associated Courses:
RADIO FREQUENCY ELECTROMAGNETIC FIELDS: environmental monitoring techniques
FUNDAMENTALS OF RADIO PROPAGATION AND ANTENNAS
UNCERTAINTIES IN MEASURING EQUIPMENT