» MEASUREMENT AND TEST INSTRUMENTATION » ELECTRIC, MAGNETIC AND ELECTROMAGNETIC FIELD METERS » ELT-400-SET-4

  

GET YOUR QUOTE

mpb datasheet mpb manuale mpb note tecniche mpb assistenza e faq

Main Category:

MEASUREMENT AND TEST INSTRUMENTATION

Secondary Category:

ELECTRIC, MAGNETIC AND ELECTROMAGNETIC FIELD METERS

Specification

Campo Elettrico: No

Campo Magnetico: Sì

Campo Elettromagnetico: No

Status:

AVAILABLE

Brand:

NARDA PMM

Model Name:

ELT-400-SET-4

Part number:

2304/104

Item Code:

NARDA PMM ELT-400-SET-4 2304/104 DB

Applications:

Measures Work Environments EMF
Environmental measures EMF LF

Description:

Set the size of the magnetic field of 1 Hz - 400 kHz with: basic unit of measurement, magnetic field sensor from 100 cm <sup>2,</sup> operating manual in English, 9 Vdc power supply. Regulatory limits set: Workers ICNIRP, IEC / EN 62233 Specification for measurements on welding

General Features:

Exposure assessment than standardized ICNIRP workers and European standard EN 50366
Operating Temperature Range: -10 ° C +50 ° C
Operating humidity range: <95% (30 ° C) or <29 g / m <sup>3</sup> without condensation
Weight: 910g (with probe)
Typical size of the base: 180 X 100 X 55 mm
Typical size of the sensor: 290 X (125 mm diameter)
Backlit LCD display
NiMH batteries (4 x AA) interchangeable
Functionality with fully charged battery: 12 hours
Charging time: 2 hours

Specifications:

Frequency range (-3dB), selectable 1 Hz - 400 kHz, 10 Hz - 400 kHz, 30 Hz - 400 kHz
Damage level sensor: 160MT (RMS) level decreases linearly with increasing frequency 77.5 Hz (1 / f)
Damage level sensor: 226mT (Peak) The level is reduced linearly with increasing frequency 626 Hz (1 / f)
Features of the peak applied for pulses of duration <= 15.6 ms and duty cycle <= 1 / 64
Typical measurement uncertainty: + / - 6% (50 Hz - 120 kHz)
Answer electric fields: <187.5 nT @ f <2 kHz, 100 V / m / <2.8 mT @ f = 50 Hz, 50 kV / m
Tripod: 1/4__
Resolution: 0.001% (low range) with standard setting
Range magnetic field: 80 mT (with EN 50366)
Resolution: 1NT (low range) with magnetic field strength approach
Analog output of the three channels X, Y, Z separated
Interface for data reading and remote control: RS232

Associated Courses:

THEORY AND MEASURES ELF (EXTREMLY LOW FREQUENCY) AND REGULATIONS
UNCERTAINTIES IN MEASURING EQUIPMENT

MORE INFORMATION

Related Articles: 3

There are no services associated with this products

GDPR COOKIE POLICY


This site uses cookies to better manage your browsing, some information will then be temporarily stored in small text files called cookies cookie.
It is very important that you are informed and accept the privacy and cookie policy of this website
For more information read our privacy and coockies policy.

privacy policy

Cookie necessari