GET YOUR QUOTE
EMI-EMC INSTRUMENTATION AND SYSTEMS
ESD GUNS
AVAILABLE
TESEQ
NSG-437
TESEQ NSG-437 STD
EMI EMC measures
Calibration for electrostatic discharge gun
Procedure according to Standard IEC 61000-4-2 Ed 2.0 (2008) and higher
DC charging voltage, up time, amplitude and pulse width after the first 30 ns to 60 ns for each level of testing (4 positive and 4 negative pulse) and 4 additional levels (2 positive and 2 negative)
Teport on measurements
Interchangeable tips for download
Powered by rechargeable batteries
Pulse according to IEC / EN 61000-4-2 (2001)
150 downloads Networks pF/330 Ω IEC / EN 61000-4-2, ISO 10605 Optional network 150 pF / 2 Ohms and 330 pF / 2 Ohms
Networks available to other Standard R = 0 Ω to 20 Ohms, C = 50 pF to 2000
LCD Display
Gun Weight 1.2 Kg; Base charging unit with battery: 5.7 Kg
Adjustable voltage from 200 V to 30 kV in steps from 100V
Measurement of discharge voltage: + / - 5%
Programmable polarity of the pulse discharge
MORE INFORMATION