If I select 3.6 GHz in set-up menu (frequency correction)when in the field there are a few sources in low frequencies ( for example FM and TV) what happens? Do I underestimate the background field
since is a broadband sensor, the correction made on a certain frequency, determines an automatic correction on the whole range. the important thing is to make a correction in frequency when the frequency that i want to correct is the prevalent one. in this way you avoid to underestimate any field.
With the probe connected to the 8053-B do I don’t see the measure on the three axes (x, y, z) What should I set?
it is normal that you do not see the 3-axis, since on this probe is not technically possible to measure the axes separately