» Frequently Asked Questions NARDA PMM   8053-B

  • The instrument won't turn on:

    - if you insert the charger plug the instrument emits a short beep but the screen is off

    - sometimes detaching and reattaching the charger plug on the screen appear a series of darker horizontal thin stripes

    - the instrument was left on charge for several hours, but it didn’t turn on anyway


    Send the product back to the service in order to replace the internal rechargeable batteries


  • I made some measurements with the avg setting, is it possible to go back to the corresponding value in rms? A clarification: since the rms value is the mean square, is always higher than the avg, am I wrong?

    the data downloaded on the PC through “ 8053 logger interface” software can then be saved in xls format. Then you can take all the individual measured values and calculate the rms average.

    Warning! This is possible only if in the set-up storage the logger setting is "1sfix. You will have all the values stored every second. It cannot be done if the setting is 6min AVG – or 1 AVG 6 min - 6.

    The RMS since is a squared value, is always a bit higher compared to AVG.


  • With ep 300 probe in plot mode I cannot see the measures scrolling on the screen

    The visualization of "Plot" indicates the variation of the field according to the time. You can't export or save the "plot" if you want to see weak variations according to the field of time, you should set the "Logarithmic" subdivision(see page 3-25 of the manual) on the main menu "SET" and on the "BAR" line (see page 3-25 of manual)


  • The storage modalities of the measure of the instrument and the cancellation of the memory of the measure on the manual it’s not very clear to me . Especially the cancellation modality

    The stored measures can be deleted only and exclusively from the instrument, overwriting above. Warning! overwriting a file implies the automatic cancellation of those in line!


  • I connected EHP-50C probe but the instrument doesn’t recognize it, on the screen appears: none probe. What should I do?

    Verifying in the SET menu if the “Optical 232” interface is selected.


  • I didn't see EHP-50C probe’s manual, where can I find it?

    It is in the 8053 manual ,in the second part.


  • Inside the box there are two optical cables apparently similar, one of them has a connector with a red protection, what is it for?

    The one with the red caps allows the connection of the instrument and / or probe directly to the PC. It has two plugs of different termination, while you can use the one with the same plugs to connect the EHP-50-C probe (or other accessories) directly to the instrument


  • From what I could see as soon as the instrument turns on the measurement starts. There is no botton to press in order to start the measurement, is it correct? What have I got to do to stop the measurement?

    As soon as the instrument is turned on , the measurement is activated, but not the recording. You can’t stop the measurement function.


  • In anticipation of high intensity magnetic field measurements in low frequency , what can I do to avoid equipment failure?

    To measure low frequencies magnetic fields(50 Hz - 100 kHz) simply hold the probe in the field and remove the instrument using the appropriate fiber optical. The EHP50C probe may be subjected to 20 mT magnetic field intensity without any problem (overload at 50 Hz) If you want to measure fields generated by MRI, EHP50 probe is not suitable for such applications where the presence of static magnetic fields of several Tesla lead to make measurements with another instrument (THM 1176) appropriate and specific for static magnetic fields.


  • what should I do if I wanted to make the measurement of SAR

    There are no user-friendly tools and affordable to calculate SAR quantities in the ICNIRP limits expressed in Dl81/08 table. Typically you use Electronic Tools (very advanced and expensive software) in order to calculate the energy specific absorption rate of mathematical human models.


ALL FAQs

GO TO THE PRODUCT